Model 204/C204
for Corrosive Liquids or Gases Gauge, Vacuum and Absolute Pressures
Applications
- High Accuracy
General Purpose
- R&D Test and
Measurement
- Vacuum Systems
- Dynamometers
- Engine Test Cells
Features
- Fast Warm-Up
- 0-5 VDC or 4-20 mA
Output
- 0.11% FS Accuracy
- Excellent
Thermal Effects
- Low Output Noise
- Fast Response, Less
than 1 Millisecond
- Solid One-Piece
Stainless Steel Sensor
- FM Approved
Intrinsically Safe
(Option)
- Meets CE Conformance
Standards
|
|
Setra’s Model 204 pressure transducers
are intended for accurate pressure
measurements of gas or liquid
media compatible with 17-4 PH stainless steel.
The high level output signal and excellent
stability, combined with fast dynamic
response, make this unit ideal for industrial,
laboratory and aerospace applications.
The exceptional accuracy is achieved by
combining Setra’s unique electronic circuitry
with its patented variable capacitance sensor.
This unit is compensated for both zero and
sensitivity shifts due to environmental temperature variations. The Model 204, with a
high-level 0-5 VDC output, and the Model
C204, with a high-level 4-20 mA output,
require no additional signal conditioning. The
sensor features a one-piece 17-4 PH stainless
steel pressure sensor and an insulated electrode,
which form a variable capacitor. As the
pressure increases, the capacitance decreases.
This change in capacitance is detected and
converted to a linear DC electric signal. On
absolute pressure units, the reference side of
the diaphragm is sealed by electron beam
welding under high vacuum.
| Model
204/C204 Specifications |
| Full Scale Pressure
Output |
Model 204
(Voltage):
5 VDC |
Model C204
(Current):
20 mA |
| Zero Pressure Output
|
0 VDC |
4 mA |
Accuracy
(RSS Method) |
±0.11% Full Scale (0.073%
Full Scale, optional, and ±0.14% Full Scale for
10,000 psig range) |
| Type of Pressure
|
Gage
Absolute
Vacuum |
| Pressure Ranges
|
0 to 25, 50, 100,
250, 500, 1000, 3000, 5000, 10000 psig
0 to 25, 50, 100, 250, 500, 1000, 3000, 5000 psia
0 to 14.7 psiv |
| Thermal Effects
|
Compensated Range °F (°C): 30 to 160 (-1 to 65)
%FS/100°F(100°C)max.zero: ±0.4 (±0.7)
%FS/100°F(100°C)max.span: ±0.3 (±0.5)
|
| Media |
Gases or liquids
compatible with 17-4PH stainless steel. (Hydrogen
not recommended for use with 17-4PH stainless
steel) |
|